Home > Producten > Gas analyzers > NDIR gas analysers > Compleet QAL 1 systeem MIR IS

Filter.

Component







































Show More


Type meting









Show More

Meettechniek































Show More

Compleet QAL 1 systeem MIR IS

In-situ Close-coupled Multi-Gas Analyzer MIR IS
A complete “all in one compact” system, for multi-gas measurements, based on the field-proven MIR 9000 analyzer and on-board SEC sampling system

Fast & simultaneous measurement of up to 10 gases among: HCl, NO, NO2 (NOx), SO2, CO, CO2, HC, CH4 (TOC), HF, N2O, O2, at the sampling location
• Integrated sample drying & system conditioning; no sample line necessary

Kenmerken en voordelen
  • Robuuste analysator met een roestvrijstalen behuizing
  • Ontworpen voor het meten van natte en corrosieve monsters
  • Eenvoudige installatie (enkele stapelinvoer, op stapel of dicht gekoppeld) voor lagere kosten
  • Flow-, temperatuur- en drukparameters (optioneel)
  • Afstandsbediening en weergavefuncties
  • Grafisch LCD-scherm met realtime synoptische gegevens
  • Intrinsieke beveiliging met ingebouwde rest-H2O-meting
  • Geïntegreerde paramagnetische cel voor langdurige O2-meting
  • Snelle responstijd, met automatische kruisinterferentiecorrectie
  • Robuuste en betrouwbare korte extractieve analysator met ingebouwd permeatiemonsterdroogsysteem voor het meten van natte en corrosieve monsters
  • Volledige CEMS in één kast - combineert infraroodanalysator, monsterconditioneringssysteem en sonde in één apparaat
  • Zeer nauwkeurig; uitstekende kalibratiestabiliteit met automatische optische controle
  • Verwarmde sonde in verschillende materialen en lengtes om aan de toepassing te voldoen
  • Externe signaalingang voor maximaal 5 extra metingen (stroomsnelheid, druk, gastemperatuur...) of een andere analoge ingang
Belangrijkste toepassingen
  • Industriële boilers en ovens, chemische en petrochemische fabrieken
  • Verbranding, warmtekrachtkoppeling, gasturbines, elektriciteitscentrales
  • Procesregeling, pre-/post-scrubbermetingen
  • Steenkool-/olie-/gasgestookte verbrandingsbewaking
  • WKK-verbrandingsbewaking
Measurable parameters
Pollutant Range
HCI 0-15 / 5000
HF 0-20 / 300
NO 0-100 / 5000
NO2 0-100 / 1000
N2O 0-20 / 1000
NOx 0-200 / 5000
SO2 0-75 / 5000
CO 0-75 / 5000
CH4 0-10 / 1000
THC 0-50 / 5000
CO2(%) 0-10 / 100
O2(%) 0-10 / 25
Lowest / Highest available ranges expressed in mg/m3 (may vary with your site conditions to be indicated on the Site Survey Form)
Technical Specifications
QAL1 to EN 15267-3 certified YES
US EPA compliance YES
Sampling technology In-Situ
Sample transfert N/A
back flush function YES
span injection YES at sample point
Built-In O2 option YES
Serial link communication RS232 / RS422
Other Communication Ethernet
Analogue OUTPUTS YES (x 8)
Analogue INPUTS YES (x 4)
Dimensions (mm) SS Tight box: 200 x 600 x 600 (D x W x H)


Goedkeuringen en certificeringen
  • COMPLIANCE WITH WID AND LCPD APPLICATIONS
  • QAL 1 certified by the MCERTs to EN 15267-3
  • MCERTs QAL 1 certified gas monitor
  • Sira MC 040030/09
  • QAL3 compliance as defined by EN 14181
  • U.S. EPA approved as compliant with 40 CFR 60 ET 75

Documentatie

envea_mir-is_insitu_multi_gas_ir_gfc_analyzer_en envea_mir-is_insitu_multi_gas_ir_gfc_analyzer_en
envea_catalogue_cems_emission-monitoring_en envea_catalogue_cems_emission-monitoring_en

Ankersmid Process
Samuel Morsestraat 4
7442 DH Nijverdal
Netherlands


(31)-0548-59 59 69
(31)-0548-62 19 72
info@ankersmid.com

© 2024 - Ankersmid Group