Meet Kwik Hg Merk: Envea Techniek: UV Industrie: Chemie Energie Luchtvaart Marine Petrochemie Type meting: Extractief Systeem

SM-4 QAL1 kwik analyser

QAL 1 gecertificeerd voor continue, nauwkeurige en betrouwbare meting van totale kwikemissies, zelfs in zeer lage concentraties en binnen complexe matrices (SO2, NOx, HCl, etc.)Totale kwikemissiemonitor in schoorsteenrookgassen, qal 1 gecertificeerd SM-4

QAL 1 gecertificeerd voor continue, nauwkeurige en betrouwbare meting van totale kwikemissies, zelfs in zeer lage concentraties en binnen complexe matrices (SO2, NOx, HCl, etc.)

Met meer dan 20 jaar ervaring was Mercury Instruments (onderdeel van ENVEA Group) de eersten die kwikmonitors en kwikanalysatoren ontwikkelden met behulp van een droge thermokatalytische methode in plaats van natte chemie.

De SM-4 Mercury Monitor is QAL1-gecertificeerd en is klaar om te voldoen aan toekomstige wettelijke vereisten (Conventie van Minamata en Europese richtlijnen voor thermische energiecentrales en verbrandingsovens).

KENMERKEN & VOORDELEN
  • Automatische en continue bewaking van de totale kwikemissies in rookgassen van schoorstenen. QAL 1 gecertificeerd
  • Extreem hoge precisie en betrouwbare meting
  • Automatische terugspoelfunctie
  • Continue meting van totaal Hg (elementair, ionisch & organisch)
  • Geen vloeibaar afval of condensaat gegenereerd
  • Monsterverdunning direct op het monsterpunt, waardoor een groot meetbereik mogelijk is en interferentie door matrixgassen wordt geëlimineerd, geen transport van ionisch Hg
  • Bemonsteringssysteem - bestaande uit de sonde, de verdunningseenheid en de lagetemperatuurkatalysator - wordt volledig verwarmd om kwikadsorptie te voorkomen
  • Zeer onderhoudsarm (service-interval: ≥ 3 maanden) – kan worden geautomatiseerd
  • Ultra hoge gevoeligheid: laag meetbereik is 0,05 µg/m³
  • Goedgekeurd en gecertificeerd door TÜV Rheinland (QAL1)
  • Belangrijkste opties:
  • Mercurius soortvorming
  • Elementaire Hg-kalibratiegasgenerator NIST traceerbaar, geïnstalleerd in de analysatorkast
  • Automatische kalibratiesystemen (geoxideerd en elementair)
  • Automatische QAL3
  • Lengte bemonsteringsleiding tot 100 m (300 ft)

BELANGRIJKSTE TOEPASSINGEN
Kwikmonitor perfect aangepast voor continue kwikanalyse in:
  • Kolencentrales (voor & na kwikabsorbers)
  • Emissiebewaking voor installaties onder 13. en 17. BImSchV
  • Cementovens
  • Zwavelzuur producerende gieterijen
  • Afvalverbrandingsinstallaties (industrieel afval, medisch afval, zuiveringsslib…)
  • Thermische behandeling van verontreinigde gronden, gevaarlijk afval
  • Metallurgische installaties met potentiële Hg-emissies...
TECHNICAL SPECIFICATIONS
Measuring principle thermocatalytic reduction at stack, Amalgamation (MI GoldTrap),
cold vapor atomic absorption (CVAAS) at 253.7 nm
Measured component Total mercury (ionic and elemental speciation as option)
UV-Source Electrodeless low pressure mercury lamp (EDL)
Stabilization method Reference beam method
Optical cell Fused silica (Suprasil), l: approx. 230mm, heated approx. 45°
Measuring ranges 0.05 μg/m³ – 500 μg/m³ Hg (up to 1000 µg/m³ as an option)
Detection limit 0.0001 μg/m³ (Detector); < 0.01 μg/m³ (System)
Response time T90 180 – 360 sec
Calibration – automatic, with built-in calibration gas generator for elementary Hg (option);
– automatic, with built-in calibration gas generator for ionic Hg (option);
– manual, with built-in heated fitting to connect externally provided calibration gas (e.g. HOVACAL®)

TECHNICAL DATA
Signal outputs analogue: 4-20 mA (500 Ω max.);
serial: RS232;
Modbus RTU/RS485 (option);
Ethernet (option)
Status output 3 x pairs of relay contacts (dry contacts)
Remote access Modem and SM-4 software (option)
Max. sample temperature 250 °C (others upon request)
Ambient temperature range allowed -5 to 40 °C
Sample line length up to 30 m (up to 100 m option)
Power supply 230 VAC / 50 Hz (110 VAC / 60 Hz option)
Electrical Power consumption Analyzer cabinet: 450 VA;
Probe-converter unit: 1000 VA;
heated sample line: 30 VA / meter
Dimensions (H x W x D) Analyzer cabinet: 161 x 80 x 60 cm;
Probe converter unit: 36 x 36 x 66 cm;
Probe controller: 48 x 44 x 27 cm
Weight Analyzer cabinet (including automatic calibrator): 130 kg;
Probe converter unit: 28 kg;
Probe controller: 14 kg

GOEDKEURINGEN & CERTIFICERINGEN

Naleving van:
  • US EPA 40CFR60 PS-12A Mercury CEMS;
  • Minamata-conventie & Europese richtlijnen voor thermische centrales en verbrandingsovens
  • EN 15267 (Europese regelgeving);
  • Richtlijn industriële emissies (IED)
  • Richtlijn geïntegreerde bestrijding van verontreiniging (IPPC),
  • Richtlijn grote stookinstallaties (LCPD)
  • Middelgrote stookinstallatie – MCP-richtlijn (EU) 2015/2193
  • Richtlijn afvalverbranding (WID)
  • Goedgekeurd en QAL 1 gecertificeerd door TÜV Rheinland Energie und Umwelt GmbH, Köln, (QAL1)

Onze kwikanalysers zijn vervaardigd volgens de ISO 9001 kwaliteitsvoorschriften.

CE goedgekeurd

Documentatie

envea_catalogue_cems_emission-monitoring_en envea_catalogue_cems_emission-monitoring_en
SM-4_Mercury_Stack_Gas_Monitor_EN_1019 SM-4_Mercury_Stack_Gas_Monitor_EN_1019
ENVEA_Process_Catalogue_EN ENVEA_Process_Catalogue_EN

Ankersmid Process
Samuel Morsestraat 4
7442 DH Nijverdal
Netherlands


(31)-0548-59 59 69
(31)-0548-62 19 72
info@ankersmid.com

© 2024 - Ankersmid Group