Modular mercury stack gas monitor SM-4 mobile

Modulaire kwik stack-gasmonitorTechnisch identiek aan de SM-4, is dit apparaat ideaal voor kwikbepaling op verschillende locaties.

De SM-4 Mobile van Mercury Instruments is ontwikkeld op basis van de beproefde SM-4 mercury stack gasmonitor. Het meetprincipe en de techniek, de thermokatalytische reactor, de detector en de monstersonde zijn identiek aan die van de stationaire SM-4.

KENMERKEN & VOORDELEN
  • Gemakkelijk te vervoeren
  • Snel gedemonteerd en weer in elkaar gezet
  • Monsterverdunning direct bij de sonde: elimineert interferentie met vrijwel elke monstermatrix
  • Onderhoudsvrije lage temperatuur converter direct bij stack: geen transport van ionisch kwik waardoor kwikadsorptie wordt geminimaliseerd
  • Detecteert totaal kwik (elementair, ionisch en gebonden kwik)
  • Automatische kalibratie voor elementair en ionisch Hg
  • Automatische kwaliteitscontrole (QAL3) tijdens bedrijf
  • Extreem breed meetbereik:
  • 0,05 tot 1000 µg/m³

BELANGRIJKSTE TOEPASSINGEN
  • Kolencentrales
  • Afvalverbrandingsinstallaties
  • Zuiveringsslibverbrandingsinstallaties
  • Cementovens
  • Registratie van Hg-emissiesituaties en bepaling van de efficiëntie van kwikreductiesystemen
  • Methode 30A toepassingen
  • Testen van stationaire CEM's
  • Optimalisatie van kwikcontroletechnologieën


TECHNICAL SPECIFICATIONS
Measuring principle thermocatalytic reduction at stack, Amalgamation (MI GoldTrap), cold vapor atomic absorption (CVAAS) at 253.7 nm
Measured component Total mercury (ionic and elemental spe
UV-Source Electrodeless low pressure mercury lamp (EDL)
Stabilization method Reference beam method
Optical cell Fused silica (Suprasil), l: approx. 230mm, heated approx. 45°
Measuring ranges 0,05 μg/m³ to 500 μg/m³ Hg
Detection limit 0,0001 μg/m³ (Detector); < 0,01 μg/m³ (System)
Response time T90 Typically 180 sec
Calibration automatic, with built in calibration gas generator for elementary Hg (option)
automatic, with built in calibration gas generator for ionic Hg (option)
manual, with built in heated fitting for feed of externally provided calibration gas (e.g. HOVACAL®)

TECHNICAL DATA
Signal outputs analogue: 4-20 mA (500 Ω max.)
serial: RS232
Modbus RTU/RS485 (option)
Ethernet (option)
Status output 3 x pairs of relay contacts (dry contacts)
Remote access Modem and SM-4 software (option)
Max. sample temperature 250 °C
Ambient temperature range allowed -5 to 40 °C
Sample line length up to 90m
Power supply 230 VAC / 50 Hz (110 VAC / 60 Hz option)
El. Power consumption Analyzer cabinet: 450 VA
Air condtioning cabinet: 1000 VA
Heated sample line: 30 VA per meter
Dimensions (H x W x D) Analyzer cabinet: 60 x 60 x 60 cm
Air conditioning cabinet: 73 x 60 x 60 cm
Probe: 75 x 38 x 34 cm
Probe controller: 48 x 44 x 27 cm
Weight Analyzer cabinet: 50 kg
Air conditioning cabinet: 50kg
Probe converter unit: 28kg;
Probe controller: 14 kg
Air consumption peak 35l/min, 6 bar (87 PSI)

GOEDKEURINGEN & CERTIFICERINGEN
De continue kwikanalysator SM-4 maakt het mogelijk om te voldoen aan toekomstige meetverplichtingen (Conventie van Minamata en de Europese richtlijnen voor thermische centrales en verbrandingsovens).

  • Naleving van US EPA 40CFR60 PS-12A Mercury CEMS;
  • Onze producten worden vervaardigd volgens de kwaliteitsnormen ISO 9001.
  • CE goedgekeurd

Documentatie

ENVEA_Process_Catalogue_EN ENVEA_Process_Catalogue_EN
SM-4_mobile_Mercury_Stack_Gas_Monitor_EN_1019 SM-4_mobile_Mercury_Stack_Gas_Monitor_EN_1019
envea_catalogue_cems_emission-monitoring_en envea_catalogue_cems_emission-monitoring_en

Ankersmid Process
Samuel Morsestraat 4
7442 DH Nijverdal
Netherlands


(31)-0548-59 59 69
(31)-0548-62 19 72
info@ankersmid.com

© 2024 - Ankersmid Group